Product was successfully added to your shopping cart.
  • Valleylab Force FXc
Previous

Valleylab Force Triad

Previous

Valleylab Force FX

Valleylab Force FXc

COVIDIEN VALLEYLAB

Quick Overview

Type: Electrosurgical Unit
Make: Valleylab
Model: Force FXc
Condition: Refurbished
Availability: In-Stock
Status: Supported by Denova Medical, Inc.

Details

Valleylab Force FX-C Description:

The Valleylab Force FX-C is an electrosurgical generator. It provides the operator with the power required for cutting, desiccating, and fulgurating tissue during both bipolar and monopolar electrosurgical procedures.  You can use handset or footswitch controls to activate the generator. Receptacles are marked to ensure the correct connection of accessories. Indicators alert you when the generator is activated. The Force FX™ allows you to adjust the activation tone volume and reset the mode and power settings used previously. An RF activation port, RS-232 serial port, and expansion port are also included. The Force FX™ can be used in conjunction with the Valleylab Force GSU System and the Valleylab Electroshield Monitor. Refer to the manuals provided with these products for more information.


Valleylab Force FX-C Specifications:

Dimensions

  • Height: 4.375"
  • Width: 14"
  • Length: 17"

 

Weight

  • 18 lbs

 

Input Power Requirements

  • Operating Range: 170 – 264.
  • VAC Max Current: 3.5 A in cut and 2 A in coag.

 

Output Waveforms

  • Bipolar - Precise, standard and Macros: 470 kHz sinusoid

 

Monopolar Cut

  • Low: 390 kHz sinusoid. Similar to the pure cut mode except the maximum voltage is limited to a lower value.
  • Pure 390 kHz sinusoid. Blend: 390 kHz bursts of sinusoid, recurring at 27 kHz intervals, 50% duty cycle envelope.

 

Monopolar Coag

  • Desiccate: 240 kHz sinusoid repeated at 39 kHz, 8% duty cycle.
  • Fulgurate: 390 kHz damped sinusoidal bursts with a repetition frequency of 57 kHz into 300Ohm.
  • Spray: 390 kHz damped sinusoid bursts with randomized repetition centered at 28 kHz.
  • Frequencies include 21 kHz < f < 35kHz.
  • Output is further modulated by a random 250 Hz envelope with a variable duty cycle.

 

Low Frequency Leakage

  • 50 – 60 Hz. Source current, patient leads, all outputs tied together.
  • Normal polarity, intact chassis ground < 10μA.
  • Normal polarity, ground open < 50μA. Reverse polarity, ground open < 50μA.
  • Sink current at high line, all points < 50μA.

 

High Frequency Leakage

  • Bipolar: < 59.2 mArms
  • Monopolar: < 150 mArms
  • CEM™: < 150 mArms at < or equal 50 W.

  Power Readouts:

  • Agree with actual power into rated load to within ±15% or 5 watt, whichever is greater.

Additional Information

Name Valleylab Force FXc
Part Number FXc, FX-C
Manufacturer COVIDIEN VALLEYLAB
Specialty Operating Room
Product Type Electrosurgical Units
Name Force FXc
Watts 300 Watts
kHz 390 kHz
Type Low, Pure, Blend
Watts 120 Watts
kHz 240 kHz
390 kHz
Type Desiccate, Fulgurate, LCF Fulgurate, Spray
Watts 70 Watts
kHz 470 kHz
Type Precise, Standard, Macro
REM Tech. Instant Response Technology, REM
in / cm 18 in / 45.7 cm
in / cm 14 in / 35.6 cm
in / cm 4.4 in / 11.1 cm
lbs / kg 18 lbs / 8.2 kg